98 . OilExpress 4 System User's Guide
6.
Disconnect the tubing to the lower port of the flow cell from the injection station
(Figure 53).
Turn the lock hub on the Luer-lok connector approximately one half-turn counter-
clockwise, until loose, and then pull to remove from the port on the injection station.
Figure 53 Disconnect Luer-lok connection at the injection station
7.
Unfasten the thumbscrew on the shuttle hinged arm securing clip (Figure 54).
Figure 54 Flow cell installed in sample shuttle
Securing clip