Overview of the OilExpress 4 System . 47
The sample shuttle is powered using the same 18 V power supply adaptor as the
spectrometer. A power splitter connected to the power adaptor port on the rear of the
Spectrum Two extends round the front of the spectrometer and connects to the front of the
sample shuttle (Figure 26).
Figure 26 Power connector on the front of the sample shuttle
When connected to the mains power, the instrument will initialize and will then be ready to
communicate with the PC. This will take approximately 2 minutes. During the initialization
procedure the diagnostic LED on the rear of the instrument is yellow indicating that the
instrument is running a series of self-checks. At the end of the initialization period, the
diagnostic LED will flash orange. The sample shuttle will also initialize, and the shuttle will
move. When the spectrometer is on the Power ON LED on the front of the instrument will be
lit green.
The spectrometer is connected via a USB connection (Figure 27) to a powered USB hub,
which is in turn connected to the PC running OilExpress 4.
Figure 27 USB 2.0 cable connector and USB port
NOTE: DO NOT connect the USB cable, or the powered hub, to the USB expansion card
installed in the PC for the (optional) Ultrasonic Liquid Level Sensor bar.
USB port
USB 2.0 cable spectrometer
connection
Summary of Contents for OilExpress 4
Page 1: ...OilExpress 4 Oil Condition Monitoring System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 17: ...Warnings and Safety Information ...
Page 27: ...Overview of the OilExpress 4 System ...
Page 67: ...Getting Started with OilExpress 4 ...
Page 95: ...Maintenance and Troubleshooting ...
Page 144: ...144 OilExpress 4 System User s Guide ...
Page 145: ...Appendices ...