Maintenance and Troubleshooting . 133
FCM – Timed out waiting for
the fluid line to achieve
cavitation (continued)
Tubing to cell or FCM has a leak
or blockage
Check connections and tighten, if
required
on page 120
If you are unable to resolve the
issue, contact your PerkinElmer
Service Representative
FCM hardware failure
Contact your PerkinElmer Service
Representative
Provide the log file for the FCM
that failed, if required
page 48
FCM – Timed out
equilibrating the waste
transfer container
Exhaust tubing is in the waste
liquid or is blocked
Remove tubing from the liquid, if
required. Empty the waste
container, if required
Check connections and tighten, if
required
FCM hardware failure
Contact your PerkinElmer Service
Representative
Provide the log file for the FCM
that failed, if required
page 48
FCM – Timed out transferring
from the waste container to
the external waste bottle
Tubing from the FCM to the
external waste container has a
leak or blockage
Check all the connections at the
FCM
Waste and/or exhaust tubing is
in the waste liquid
Remove tubing from the liquid
Empty the waste container, if
required
FCM hardware failure
Contact your PerkinElmer Service
Representative
Provide the log file for the FCM
that failed, if required
page 48
DUO system operates as a
single-spectrometer system
only
Both spectrometers are not
setup on the Spectrometer Setup
dialog
Confirm that the correct
spectrometers are selected on the
Spectrometer Setup dialog
Summary of Contents for OilExpress 4
Page 1: ...OilExpress 4 Oil Condition Monitoring System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 17: ...Warnings and Safety Information ...
Page 27: ...Overview of the OilExpress 4 System ...
Page 67: ...Getting Started with OilExpress 4 ...
Page 95: ...Maintenance and Troubleshooting ...
Page 144: ...144 OilExpress 4 System User s Guide ...
Page 145: ...Appendices ...