22 . OilExpress 4 System User's Guide
Mechanical Hazards
WARNING
When the autosampler is in operation, mechanical parts may move
suddenly, and without warning. The autosampler is designed for
automatic, hands-off operation only.
NEVER reach into the autosampler workspace when the autosampler is in
operation!
Environment
To obtain the best performance from your OilExpress 4 System:
•
The laboratory must be well-ventilated, relatively dust-free, and equipped with
extraction to remove any build up of solvent vapor.
•
DO NOT place the OilExpress 4 System near to room heating equipment, for example,
central-heating radiators or cooling units, in direct sunlight, or where it will be exposed
to strong magnetic fields.
•
Ensure that there are no overhanging shelves, and no water pipes or faucets that could
leak onto the OilExpress 4 System.
•
The OilExpress 4 system must be installed on a flat and level bench (such as
part number 5085685), free from vibrations or mechanical shocks.
•
The bench must be strong enough to support the weights specified in Table 4.
Table 4 Dimensions of the OilExpress 4 System
System
Dimensions*
Width (mm)
Depth (mm)
Height (mm)
Weight (kg)
OilExpress 4
1700
830
850
149
OilExpress 4 DUO
1700
830
850
169
OilExpress 4 XL
2210
830
850
174
OilExpress 4 XL DUO
2210
830
850
194
* The widths and weights dimensions detailed in the table above include the PC and LCD monitor.
The depth values above are for the autosampler with closed safety shield. Opening the shield
requires a total depth clearance of 1080 mm.
Summary of Contents for OilExpress 4
Page 1: ...OilExpress 4 Oil Condition Monitoring System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 17: ...Warnings and Safety Information ...
Page 27: ...Overview of the OilExpress 4 System ...
Page 67: ...Getting Started with OilExpress 4 ...
Page 95: ...Maintenance and Troubleshooting ...
Page 144: ...144 OilExpress 4 System User s Guide ...
Page 145: ...Appendices ...