130 . OilExpress 4 System User's Guide
FCM fails fit-for-purpose test
(continued)
FCM hardware failure
Contact your PerkinElmer Service
Representative
Provide the log file for the FCM
that failed, if required
page 48
Valve or Pump failure
–
Possible short circuit or
overcurrent
FCM hardware failure
Contact your PerkinElmer Service
Representative
Ultrasonic Liquid Level
Sensor Bar fails to calibrate
Items on deck other than
allowed labware
Check that the labware is located
in the correct positions for the
current configuration
Clear the deck of any obstruction
Items off deck, to the left of the
calibration bar
Remove any items
Sensor bar hardware failure
Contact your PerkinElmer Service
Representative
Ultrasonic Liquid Level
Sensor Bar fails to detect
heights
Sample level too low
Check that all sample levels are
above the minimum height for the
sample container
Sample not present
Ensure that all containers contain
samples and are loaded into the
sample racks in the correct
sampling order
Sensor bar is not connected to
the PC (via additional card)
Ensure that the sensor bar is
connected to the USB expansion
card fitted to the PC
Sensor bar hardware failure
Contact your PerkinElmer Service
Representative
System liquid tubing contains
air after priming
System liquid tubing is not
immersed in system liquid
Check the tubing is immersed in the
system liquid. Refill the system
liquid container, if required
Use the Prime button on the
Autosampler Control dialog to
Prime system 2–3 times more (until
there is no air in the tubing) before
on page 74
Summary of Contents for OilExpress 4
Page 1: ...OilExpress 4 Oil Condition Monitoring System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 17: ...Warnings and Safety Information ...
Page 27: ...Overview of the OilExpress 4 System ...
Page 67: ...Getting Started with OilExpress 4 ...
Page 95: ...Maintenance and Troubleshooting ...
Page 144: ...144 OilExpress 4 System User s Guide ...
Page 145: ...Appendices ...