Warnings and Safety Information . 19
OilExpress 4 System Safety Summary
This equipment must be used only by authorized personnel in a well-ventilated laboratory
environment.
The PerkinElmer OilExpress 4 System has been designed to comply with a wide variety of
international standards governing the safety of laboratory equipment. Refer to the product
certificates and declaration of conformity documents supplied with the components of your
OilExpress 4 System for the relevant compliance information.
In routine use, the OilExpress 4 System poses virtually no risk to you. If you take some
simple, common-sense precautions, you can make sure that you maintain the continued safe
operation of the OilExpress 4 System.
•
DO make sure that the OilExpress 4 System is properly connected to the electrical
supply; in particular make sure that the ground (earth) is securely connected.
•
DO take care with solvents: heptane is highly flammable. Fill the system liquid container
away from the OilExpress 4 System and any heat sources or other electrical systems.
Always keep the lid on the system liquid and waste liquid containers.
•
The system waste liquid container must be vented externally or connected to an
extraction system.
A connector for 6 mm (internal diameter) tubing is fitted to the lid supplied with the
waste container for the OilExpress 4 System.
•
DO NOT place any part of the body beyond the safety shield the front of the
autosampler enclosure whilst operating the OilExpress 4 System.
•
DO disconnect the electrical power cable before opening any ports or covers in the
system, unless otherwise directed in the maintenance procedures described in this
manual or in the software.
•
DO NOT open the main cover of any of the components of the instrument, unless
directed in the maintenance procedures.
•
DO keep the spectrometer dry. Avoid spilling liquid onto the instrument. Clean all
external spills immediately.
Read the more detailed information about warnings and safety in the following pages, and in
the documentation supplied with your spectrometer, to ensure the safe operation of the
instrument.
Summary of Contents for OilExpress 4
Page 1: ...OilExpress 4 Oil Condition Monitoring System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 17: ...Warnings and Safety Information ...
Page 27: ...Overview of the OilExpress 4 System ...
Page 67: ...Getting Started with OilExpress 4 ...
Page 95: ...Maintenance and Troubleshooting ...
Page 144: ...144 OilExpress 4 System User s Guide ...
Page 145: ...Appendices ...