122 . OilExpress 4 System User's Guide
Changing to a Single-spectrometer System (DUO
Systems Only)
If you have DUO system and one of your spectrometers or FCM(s) requires maintenance, you
can switch to a single-spectrometer system to continue analyzing samples.
1.
Select
Spectrometer
from the Instrument sub-menu of the Setup menu.
The Spectrometer Setup dialog is displayed.
2.
Deselect the appropriate spectrometer,
Lower
or
Upper
.
If the FCM has failed, you must ensure that you disable the spectrometer at the same
level in the stacking system.
3.
Disconnect the spectrometer and corresponding FCM from the mains power supply.
Removing the Sample Shuttle
Ensure that you wear appropriate protective gloves when handling fluid lines that may
contain harmful materials.
To remove the sample shuttle from the sample compartment of the spectrometer:
1.
Ensure that the flow cell is empty.
If required use the Flush options on the Autosampler Control screen to flush and dry the
flow cell.
2.
Exit OilExpress 4 software.
3.
Disconnect the spectrometer and FCM from the mains power supply.
If you have a DUO system, this should be the FCM connected to the appropriate
spectrometer.
4.
Disconnect the liquid presence sensor from the rear of the FCM.
5.
Disconnect the power cable from the front of the sample shuttle.
Summary of Contents for OilExpress 4
Page 1: ...OilExpress 4 Oil Condition Monitoring System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 17: ...Warnings and Safety Information ...
Page 27: ...Overview of the OilExpress 4 System ...
Page 67: ...Getting Started with OilExpress 4 ...
Page 95: ...Maintenance and Troubleshooting ...
Page 144: ...144 OilExpress 4 System User s Guide ...
Page 145: ...Appendices ...