Overview of the OilExpress 4 System . 51
Fluid connections on the FCM
The fluid connections on the rear of the FCM are shown in Figure 31 (shown with the
connectors installed in the mounting block). Fluid tubing connections on the rear of the FCM
connect the upper (outlet) port on the flow cell to the waste transfer container (WTC) in the
FCM, and the WTC to the external waste liquid container.
The EXHAUST outlet vents solvent vapor from the pump to the waste container. Ensure that
the waste liquid container is lower than the FCM so that any vapor condensing in the line will
drain to the container. The waste liquid container should be vented to a suitable outlet.
Figure 31 Fluid connections on the FCM
CELL (from the flow cell)
EXHAUST (to waste container)
WASTE (to waste container)