Overview of the OilExpress 4 System . 65
17.
The autosampler arm moves the disposable tips to the first available empty waste tip
location, and the two used tips are dropped into positions 1 and 3.
18.
The autosampler moves to the first available clean tip and loads a disposable tip on
VersaTips 1 and 3 from the next two available positions.
These are collected from positions 2 and 4.
19.
The autosampler returns to the injection station and positions VersaTips 2 and 4 over
Fill Cups 1 and 2.
20.
When the spectrum has been collected, the FCM draws the sample to its internal waste
container.
The spectrometer status is
Idle
. The liquid presence sensor LED changes from green to
red.