140 . OilExpress 4 System User's Guide
Empty Cell Spectrum macro
fails repeatedly
Refer to the message box for the
cause of the failure
Follow the suggestions on the
on page 105
If, after taking any appropriate
action, the Empty Cell Spectrum
macro continues to fail, then
contact your PerkinElmer Service
Representative
Pathlength Calibration fails
repeatedly
page 109 for possible causes
on page 109 for
possible solutions
If, after taking any appropriate
action, the Pathlength Calibration
continues to fail, then contact your
PerkinElmer Service Representative
AVI Calibration fails
Beam is partially or completely
blocked
Check the sample compartment
and ensure that the beam path is
clear. Repeat the calibration
Spectrometer hardware failure
Contact your PerkinElmer Service
Representative
If any of these problems persist, or if there are any other problems that you cannot correct,
contact your local PerkinElmer office or agent.