74 . OilExpress 4 System User's Guide
Priming the System
The system liquid in the fluid lines functions as the hydraulic fluid for the system. Air in the
lines can affect accurate dispensing of samples and flushing. Refer to
on page 148 for more information about liquid handling in the OilExpress 4 System.
We recommend that you prime the system once a day (before collecting data), or whenever
you observe air in the fluid lines.
Each time you start an analysis (a batch, system check or pathlength calibration) an
Initialization dialog is displayed (Figure 43), which includes the option
Prime autosampler
on initialization
. Inspect the syringes and tubing to the VersaTips for any air bubbles and, if
required, select the option to prime the system.
NOTE: This option will be selected by default the first time an analysis is run after starting
the OilExpress 4 software application.
Figure 43 Batch Initialization dialog showing Prime option
During the prime, each syringe is twice filled with system liquid, which is dispensed via the fill
cup through the fluid lines. Each syringe is then filled with a smaller volume of system liquid,
which is used to fill the injection station fill cup. The corresponding VersaTip is then lowered
into the fill cup and cleaned. This procedure takes approximately two minutes in total.
NOTE: During priming, check all tubing, syringes and connections for leaks, particularly after
any maintenance procedures. Persistent air bubbles, or system liquid dripping from
tips, are indicative of leaks.
Summary of Contents for OilExpress 4
Page 1: ...OilExpress 4 Oil Condition Monitoring System User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 17: ...Warnings and Safety Information ...
Page 27: ...Overview of the OilExpress 4 System ...
Page 67: ...Getting Started with OilExpress 4 ...
Page 95: ...Maintenance and Troubleshooting ...
Page 144: ...144 OilExpress 4 System User s Guide ...
Page 145: ...Appendices ...