Overview of the OilExpress 4 System . 31
For FT-IR analysis using OilExpress 4 software, two of the VersaTips are dedicated to
aspirating samples for analysis using wide-bore disposable tips (VersaTips 1 and 3). The
other two are dedicated to dispensing system liquid, for example to clean the flow cell
Figure 6 VersaTips 1 and 3 (with disposable tips) are used for sampling;
VersaTips 2 and 4 are used for dispensing system liquid
For systems running in the dual FT-IR analysis and oil dilution mode, VersaTips 1, 2, 3 and 4
are used with disposable tips for oil sampling. VersaTips 1 and 3 are used without tips to
dispense solvent for dilution. VersaTips 2 and 4 will be used to dispense system liquid to
clean the flow cell.
System Liquid Container(s)
WARNING
In the event of a solvent spillage, isolate the power to the OilExpress 4
System components immediately and then clean up the solvent spill.
The system liquid container must be placed next to the autosampler in a location where it
will not be knocked over.
NOTE: The system liquid container available with the OilExpress 4 system is 2 L
(part number L9004516) and the waste container is 5 L (part number L9004517).