Maintenance and Troubleshooting . 117
14.
Screw the blue nut connector from the tubing from the lower port of the flow cell into a
red Luer-lok connector.
15.
Connect the tubing from the lower port of the flow cell to the appropriate fill cup on the
injection station.
For a single-spectrometer system, the tubing is connected to Fill Cup 1 (shown in
Figure 67), which is to the rear of the autosampler deck. For a DUO (dual-spectrometer)
system, lower flow cell is connected to Fill Cup 1, and the upper flow cell is connected
to Fill Cup 2.
Push the Luer-lok connector into the connection on the injection station, then turn the
lock hub on the Luer-lok connector approximately one half-turn clockwise Figure 68).
Figure 67 Connecting tubing to Fill Cup 1
Figure 68 Tubing connected to Fill Cup 1