64 . OilExpress 4 System User's Guide
12.
The disposable tips are lowered into the fill cups to form a seal and the syringe pump
dispenses the sample to the fill cup.
13.
The liquid presence sensors detect the presence of the samples in the tubing to the flow
cell.
Each liquid presence sensor detects the presence of the front of the liquid moving
through the tubing from the fill cup, and sensor LED changes from red to green. The
time taken for the liquid to reach the sensor is used by the OilExpress 4 software to
calculate parameters, such as the time required for the FCM to pump the sample into
the flow cell and the adaptive cleaning settings.
Only 700 µl of sample is dispensed to the cell.
14.
When the sample is detected, if required the FCM draws the sample into the flow cell.
15.
The sample spectra are collected.
The autosampler status is
Preparing
. The spectrometer status is
Scanning
.
16.
During the sample measurement, the autosampler moves the disposable tips over the
waste trough on the injection station, and any excess oil is ejected.
NOTE: For some low-viscosity samples, the VersaTips may remain in the fill cups until after
the sample scan has been completed.