3.1 System
BioFlo 415 is a versatile sterilizable-in-place (SIP) fermentor that provides a fully
equipped system in one compact package. It can be employed for batch or continuous
culture with process control of pH, dissolved oxygen (DO), agitation, temperature,
pump feed, antifoam, level and addition analog/digital inputs and outputs.
3.2 Vessels
The fermentation vessels are designed for total volumes of 7, 14 and 19.5 liters.
Each stainless steel vessel consists of a headplate, view window and magnetic drive
agitation with Rushton impellers. Ports are provided in the headplate for the following
purposes: inoculation; base and acid addition; a thermowell for a resistance
temperature detector (RTD); a foam probe; a sparger; a harvest tube; a sampling tube;
an exhaust condenser; and dissolved oxygen (DO) and pH electrodes. The drive
bearing housing is also located on the headplate
(see Figures 3a, 8, 9 and 10)
.
3.3 Agitation
System
A removable agitation motor located on top of the bearing housing on the headplate is
connected to the agitation shaft with a magnetic drive coupling.
The motor will provide a speed range from 50 to 1000 RPM. The process-control
software ensures agitation speed control from
±
1 RPM to
±
5 RPM throughout the
speed range.
It is possible to cascade Dissolved Oxygen (DO) to Agitation (AGIT) so the agitation
speed will vary between the user-specified minimum and maximum setpoints in order
to maintain the set percentage of DO.
(See Section 11 for further information on
setting up cascades.)
Default P & I (proportional & integral) values are preset at the factory.
We strongly
recommend that you maintain the factory-set parameters.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...