Table 4: AFS/Modbus Com Port Pin Designation
Pin Number
Signal
Comments
1, 4-6, 8-11,
14-20, 22-23
NC not
assigned
2
TXD
RS232 Data Output from fermentor
3
RXD
RS232 Data Input to fermentor
7
GND
Ground reference for all signals
12 IRXD+
24 IRXD-
RS422 paired data input to fermentor
13 ITXD+
25 ITXD-
RS422 paired data output from fermentor
21 IOS
Open selects RS232
Grounded selects RS422
Unless otherwise requested, the baud rate is factory-selected at 9600 (AFS) or 19200
(Modbus) and the connector is configured as an RS232 port: i.e., no jumper between
pin #7 and pin #21. The factory-set address for the machine is
8
.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...