Table 5: Changing Sterilization Parameters
Variable
How to Change
Drain Time
Press the DRAIN edit box to input the duration of the Drain phase
(0-240 minutes).
Heat B
Temperature
Press the HEAT B edit box to input the temperature at which the
vessel will transition from Heat A phase into Heat B phase. Must
be
≤
Sterilization Temperature.
See important
NOTE
below
.
Sterilization
Temperature
Press the STERIL TEMP edit box to change the temperature at
which the vessel will be held during the Sterilization sequence
(121.5 – 130.0° C).
Sterilization
(Hold) Time
Press the STERIL TIME edit box to change the amount of time
the vessel will be held at the Sterilization Temperature. (0-600
minutes; factory-set to 45 minutes)
Cool B
Temperature
Press the COOL B edit box to input the temperature at which the
vessel will transition from Cool A into Cool B phase.
See
important
NOTE
below
.
Growth
Temperature
Press the GROWTH TEMP edit box to define the temperature at
which the vessel will transition into the Growth phase.
#
NOTE:
Entering a Heat B temperature below 100
o
C will cause excessive heating
time.
#
NOTE:
Cool B temperature must be set higher than Growth temperature.
You can reset any of these parameters by pressing on the corresponding text
box and entering the desired value on the touchpad that appears. Press
OK
to
accept the parameters and to close the touchpad.
8.3 Valve
Sequencing
The controller has been preconfigured to sequence a set of valves on the BioFlo 415
during the automatic sterilization cycle. The column headings or operations for valve
sequencing are as explained in Table 6 on the following page:
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...