85
New Brunswick Scientific
User’s Guide
6.
Repeat Steps 4 & 5 for any other alarms recorded for this loop.
7.
Press the
Scroll Back
(
<<<
) button to return to the
ALARM
screen.
8.
Repeat Steps 2-7 for any other control loop alarms.
To acknowledge
all alarms simultaneously for one control loop
:
1.
Press the
ALARMS
button to open the
ALARM
screen.
2.
Press the red
LoopName
box to open that control loop’s
ALARM
screen.
3.
Press the
Current Alarms
button to open the
Current Alarms Summary
screen
(
see Figure 49
).
4.
Press the
Acknowledge All
button. All alarms will be deleted from this screen.
5.
Press the
Scroll Back
(
<<<
) button to return to the
ALARM
screen.
To acknowledge
all alarms for all control loops
at the same time:
1.
Press the
ALARMS
button to open the
ALARM
screen.
2.
Press the
Acknowledge All
button. All alarms will be deleted from this screen.
3.
Press the
Scroll Back
(
<<<
) button to return to the
ALARM
screen.
CAUTION!
Acknowledging alarms is NOT a replacement for correcting the condition
that triggered the alarm. Diagnose the cause of the alarm condition and
rectify the situation to ensure proper operation of your BioFlo 415.
14.4 Alarms
History
Each time an alarm is triggered, whether Visible and/or Audible, the controller records
the event. The controller also records each alarm acknowledgement. You can access
the Alarms History screen (1) to consult the data, (2) to save the data to an optional
auxiliary PC, and/or (3) to purge the records once the condition has been rectified.
To access the Alarms History screen (
see Figure 50
)
to consult data
:
1.
Press the desired control loop’s
LoopName
box in the main
ALARM
screen.
2.
In the control loop’s
Alarms Screen
that opens, press the
Alarm History
button.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...