Preventive maintenance keeps your equipment in proper working condition. When performed
routinely, maintenance results in longer life for your equipment. It also reduces time lost due
to equipment failure.
WARNING!
Always turn your BioFlo 415 off and disconnect the power cord before
performing maintenance.
18.1
pH Probe Maintenance and Storage
The pH probe should be stored standing upright, with the electrode tip immersed in a
solution of 3 molar KCl or a buffer solution between pH 4.00 and pH 7.00.
CAUTION!
Never let a pH probe rest on its tip.
Never leave a pH probe in DI water.
18.2
DO Probe Maintenance and Storage
Use soft facial tissue to clean the DO probe.
Check the probe’s Teflon membrane to be sure there are no punctures, puckers or
wrinkles. If there are, the probe should be replaced.
When it is not in use in the vessel, the DO probe should be stored standing upright
with the shorting cap in place and the membrane isolated from the air environment.
At no time should the probe be allowed to rest on its membrane.
CAUTION!
Never let a DO probe rest on its tip.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...