Figure 17: Calibration Screen (pH)
9.
Rinse the pH electrode with distilled water.
10.
Immerse pH electrode into a second pH buffer solution which is several pH units
above or below pH 7.00 (e.g., pH 4.00) and allow a few minutes for the system to
equilibrate.
11.
Similar to step 7 above, touch the
SET SPAN
edit box. Use the touchpad that
opens to enter the value of the second buffer solution (e.g., 4.00), then press the
OK
button.
12.
When the
CURRENT VALUE
reading stabilizes, press the
SET SPAN
button.
13.
To ensure accuracy, repeat Steps 4-11 a few times, using the same two buffer
solutions.
#
NOTE:
The pH calibration should be checked after sterilization, immediately prior
to inoculation. Take a sample from the vessel and compare the pH value
displayed on the control cabinet screen to the pH recorded by an external
pH meter. Any discrepancy should be adjusted with the
SET ZERO
procedure.
6a. Press
pH here...
6b. ...and pH
appears here.
7. Touch the SET
ZERO edit box. Use
the touchpad that
opens to enter 7.0,
then press the
touchpad’s OK
button.
8. When the
CURRENT VALUE
reading stabilizes,
press the
SET ZERO button.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...