6.
Enter a temperature setpoint that is at least 12°C below the current value.
The controller will respond to the call for cooling by opening solenoid
valve TY-5K, filling the jacket with water. temperature control mode to
OFF
.
8.2.2
Sterilization Safety Features
The heater has a safety switch that will automatically turn the heater off at
543.3ºC (650ºF).
During the sterilization cycle, if the pressure in the vessel or vessel jacket
increases, the rupture disk is set to go off at 40 PSIG and the safety release
valve on the vessel jacket is set to go off at 50 PSIG.
8.2.3
Starting a Sterilization Sequence
1.
At the touchscreen, set the Temperature control loop for the desired
operating temperature after sterilization. Turn the Temperature control
loop
ON
by setting the temperature control loop to
AUTO
control mode.
2.
Set the sparge gas flow rate at the desired setpoint for operation after
sterilization. The
GasFlo
control loop should be set to the
AUTO
control
mode.
#
NOTE
:
If you are using Gas Mixing, the O
2
loop must be in the O2ENRH mode.
3.
Ensure all headplate ports and upper side-entry ports are closed. All ports
should either have a piece of equipment inserted or be plugged. All clamps
on ports should be secure.
4.
Set the agitation rate at the touchscreen. During sterilization, Agitation
control loop should be set to a rate of 150 - 300 RPM. Turn on the
Agitation control loop by putting the loop in
AUTO
control.
5.
Press the
Sterilize
button to open the
STERILIZATION
screen (
see Figure
22, repeated on the following page for easy reference
).
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...