24
BioFlo 415
y
M1360-0050
User’s Guide
5
5
S
S
P
P
E
E
C
C
I
I
F
F
I
I
C
C
A
A
T
T
I
I
O
O
N
N
S
S
BioFlo 415 System
Total Volume
7L 14L 19.5L
Working Volume
2 - 5.25L
4 – 10.5L
5 – 15.5L
Vessels
Aspect Ratio
2:1 2:1 3:1
Control Station
Controls 1 vessel with 32 control loops; stores 10
recipes & 8 process variables per vessel for trend
graphing. Includes a touchscreen monitor/user
interface, 3 built-in pumps & connectors for all utilities
& communications signals used by the fermentor.
Controller
Touchscreen
Interface/Display
15-inch monitor capable of supporting one fermentor.
Indication
Digital display in 0.1°C increments
Range
From 5°C above coolant temperature to 80°C (setting
range: 4-80°C).
Control
PI control employing PWM of heater and cooling water
Temperature
Sensor
Platinum RTD probe
Drive
Permanent magnet motor with high torque input.
Indication
Digital display in 1 RPM increments.
Range
50-1000 RPM
Control
±1%
Bearing Housing
Standard single mechanical seal with magnetic drive
Sensor
Optical photoplastic disc 500 lines/rev with quadrature
output.
Agitation
Impellers
Six-bladed Rushton turbine impellers provided
Filter
1.2
μ
m depth filter; 0.2
μ
m absolute filter optional
Exhaust
Condenser
Stainless steel
4-Gas System
Up to 4 gases, including air, N
2
, CO
2
& O
2
, delivered to
ring sparger
Range
0.5-25 SLPM @ 20 PSIG
Sparger
Ring sparger
Inlet Filter
0.2
μ
m absolute filter
Aeration
N
2
Gas
For calibration of DO probe
Indication
Digital display in 0.01 pH increments
Range
2-12 pH
Control
P&I
pH
Sensor
pH gel-filled probe
Indication
Digital display in 0.1% increments
Range
0-200%
Control
P&I, Agitation, O
2
Enrichment. Also GasFlow Rate if
equipped with mass flow controller
DO
Sensor
Polargraphic probe
…continued…
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...