22.1
Determining an Oxygen Transfer Rate
The oxygen transfer rate (OTR) of all NBS fermentors is determined by a standard
sulfite oxidation test.
The standard operating conditions for determining OTR are:
Temperature: 30
°
C
Agitation: 1000
RPM
Aeration: 1
VVM
22.1.1 OTR
Calculations
OTR can be estimated by titrating a fixed amount of sodium sulfite, Na
2
SO
3
,
with air, CU+2:
2SO
3
+ 0
2
→
2SO
4
The Procedure
Calibrate the DO electrode:
•
Set zero on DO.
Fully oxygenate the fermentor with agitation and airflow.
•
Set span to 100%.
Introduce a known amount of Na
2
SO
3
into the fermentor when fully
oxygenated.
•
OTR = 30,000 n mM O
2
/L/hr
V
Δ
T
n
= number of moles of sodium sulfite
V
= vessel volume in liters
Δ
T
= time taken from DO curve at two points of 50% DO min.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...