7.1
pH Probe Inspection
Inspect probe for possible shipping damage. If damage is observed, notify the New
Brunswick Scientific Service Department immediately.
If you have a liquid-filled probe, check the level of the reference electrolyte. It should
be about 1cm below the filling orifice. To add reference electrolyte, take the filling
pipette (P0740-4820) and fill it with Viscolyte B (P0860-0130) electrolyte.
Check the electrode tip for trapped air bubbles. To remove any air bubbles, hold the
electrode upright and shake gently.
NEVER REST THE PROBE ON ITS TIP.
#
NOTE:
Both chambers of the electrode are filled with the same reference
electrolyte, for a total volume of approximately 30 mL. For normal
operation, remove the rubber T stoppers if the probe is so equipped,
saving them for use during sterilization.
7.2
pH Probe Calibration
#
NOTE:
Calibrate the pH probe before sterilizing it in the vessel.
1.
If you have not already done so, connect the pH probe to the pH connector on the
control cabinet (
see Figure 1a
), using the appropriate cable.
2.
Turn the main power switch
ON
.
3.
Press the
CALIBRATION
button to display the
CALIBRATION
screen.
#
NOTE:
The pH probe is calibrated using two external buffer solutions of known
pH, usually 7.00 and 4.00.
4.
Rinse the pH electrode with distilled water, then immerse it into pH 7.00 buffer
solution and allow a few minutes for the system to equilibrate.
5.
Open the
CALIBRATION
screen (
see Figure 17
, repeated below for easy
reference). Steps 6-8 are indicated as callouts around this screen:
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...