14.1
ABS and DEV Alarms
There are two types of alarm modes you can set, Absolute (ABS) and Deviation
(DEV):
An Absolute alarm is triggered when the control loop’s Process Variable falls
below the absolute Low limit or rises above the absolute High limit that you
set.
A Deviation alarm is triggered when the control loop’s Process Variable falls
below or rises above the control band that you specify around the loop’s
setpoint (e.g., a tolerance of 10 rpm above or 5 rpm below the Agitation
setpoint).
14.2 Setting
Alarms
To set alarms:
1.
Press the
ALARMS
button to open the
ALARM
screen (
see Figure 21
,
repeated here for easy reference,
and Table 10
, which explains the features
of this screen).
Figure 21: Alarm Summary Screen
2. Press the first loop
for which you want to
enable an alarm. That
loop’s individual Alarms
Screen will open. For
this example, we use
the Agitation loop.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...