Figure 42: Standard Pump Array
Before you insert tubing into the
PUMP CHANNEL
, verify that the
PUMP
is in the
OFF
control mode.
With reference to Figure 60 below
, follow these steps to properly load
tubing into the
PUMP HEAD
:
Figure 60: Loading Pump Tubing
1.
Open the
PUMP
cover to gain access to the interior of the pump.
2.
Select the desired tubing size (
see Table 9 in Section 13.3 for reference
) and cut a
length sufficient to reach from the inlet source, through the pump, and to the outlet
recipient, allowing a few extra inches.
3.
Form a loop large enough to go around the pump head.
4.
Pull the right side of the spring-loaded lever down to release the tubing guide.
5.
Hold the upper spring-loaded clip open and load the upper clip channel; allow the
upper clip to close over the tubing.
SPRING-LOADED
LEVER
TUBING GUIDE
PUMP HEAD
UPPER
SPRING-LOADED
CLIP
LOWER
SPRING-LOADED
CLIP
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...