57
New Brunswick Scientific
User’s Guide
Table 6: Valve Sequencing Labels/Operations
LABEL/OPERATION
EXPLANATION
Device
Valve or component name/ID
Growth
Cooling with temperature within 1ºC above normal
operation temperature setpoint.
Drain
Jacket priming operation
Heat A
Heating at temperatures below the Heat B
temperature setting at which steam is not generated
in the vessel.
Heat B
Heating at temperatures above the Heat B
temperature setting at which steam is generated in
the vessel.
Steril
Heating operation at sterilization temperature
Cool A
Cooling with temperature > 10ºC above normal
operation temperature setpoint.
Cool B
Cooling with temperature
≤
10ºC above normal
operation temperature setpoint.
The factory settings for the sterilization sequence are indicated in Figure 32, which is a
detail from the
STERILIZATION
screen:
Figure 32: Sterilization Sequence Settings
The legend for these settings is shown in Figure 33 on the following page, which is
also a detail from the
STERILIZATION
screen:
The Heater,
valve TY-5K,
and Exhaust
Heater
Growth
settings are
black & white.
The Heater
Steril
setting is red &
white.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...