66
BioFlo 415
y
M1360-0050
User’s Guide
10.6.5 Exhaust
Heater
The BioFlo 415 has been provided with a heater band on the exhaust line. The
purpose of the heater is to prevent the exhaust filter from becoming saturated
with condensate. To activate the heater, do the following:
If 1 TMFC is installed:
1.
Go to the
CASCADE
screen and select the GasFlo loop as the Cascade
From loop.
2.
Enable the default cascade to the exhaust heater.
3.
Go to the
SUMMARY
or
SYNOPTIC
screen and select the ExhstHtr loop.
Set it in the
MANUAL
control mode.
For all other set-ups:
1.
Go to the
SUMMARY
or
SYNOPTIC
screen and select the ExhstHtr loop.
Enter a setpoint of 85%.
2.
Set it in the
MANUAL
control mode.
WARNING!
Never turn the heater off during the run or the exhaust filter may clog.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...