89
New Brunswick Scientific
User’s Guide
15.1.1
Gas Control with 1 or No TMFC
If your system is equipped with no TMFC or one TMFC, you have the choice
of two
Operating Modes
in the
Controller Setup
screen: O2 Enrich and 3-Gas
in Fermentation mode (
see Figure 51
). Your system has 4 gas solenoid valves.
No TMFC means that all gas flow is manually controlled using a rotameter.
Figure 51: Controller Setup Screen (0-1 TMFC)
If you select O2 Enrich as the
Operating Mode
, the gas process loops you will
find in the
SUMMARY
screen are labeled Air (1) and O2 (2). If you select
3Gas, the process loops are labeled Air (1), O2 (2) and CO2 (4). The loops’
numbers, 1, 2 & 4, correspond to the gas connections on the cabinet (
see
Figure 1a
).
There is also a GasFlo loop when one TMFC is present; settings in this loop’s
gauge screen turn the TMFC on and off and control the gas flow rate.
15.1.2
Gas Control with 2 TMFCs
If your system is equipped with two TMFCs, the only
Operating Mode
available in the
Controller Setup
screen is O2 Enrich, because only two gases
can be brought into the unit (
see Figure 52 on the following page
).
The gas process loops you will find in the
SUMMARY
screen are labeled
AirFlo (1) and O2Flo (2). The loops’ numbers, 1-2, correspond to the gas
connections on the cabinet (
see Figure 1a
).
Select
O2
Enrich
for
Air and O
2
.
Select
3Gas
for Air, O
2
&
CO
2
.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...