114
BioFlo 415
y
M1360-0050
User’s Guide
Problem
Possible Solution
AGITATION:
Agitator does not turn, or turns
only slowly.
•
Make sure the motor is plugged into the cabinet
receptacle; TURN OFF SUMMARY POWER BEFORE
CONNECTING THE PLUG.
DO and pH PROBES:
DO probe readings are erratic.
•
Recalibrate the probe, carefully following instructions
in this manual.
•
Recharge the probe, carefully following instructions in
this manual.
•
Probe may need a new membrane and a refill of
electrolyte.
•
Check for a secure connection.
•
Replace probe cable or DO probe.
pH probe readings are erratic.
•
Recalibrate the probe, carefully following instructions
in this manual.
•
Check for a secure connection.
•
Gel-filled probe may need replacement.
•
Liquid-filled probe may need a refill of electrolyte.
•
Probe cable may need replacement.
Probe does not hold calibration.
•
Probe may be defective; replace it.
•
pH/DO amplifier on superboard may be defective; call
for service.
GASFLOW:
There is insufficient gas flow.
•
Inlet or exhaust sterile air filter may be wet or clogged;
replace filter.
•
Check that the air pressure is within the specified
range.
•
Make sure the control mode for DO and for pH is set
to
AUTO
(not
OFF
).
•
Make sure that the GasFlow loop is ON.
•
Make sure that the Air loop is in O2 Enrichment mode.
•
Make sure that the DO cascades are Enabled.
Gas is entering vessel but
display is not reading gasflow.
•
Mass Flow Controller has been penetrated by
moisture or particulate matter. Contact NBS for
repair/replacement.
STERILIZATION:
Unit does not reach sterilization
temperature
•
Water pressure is too high (
>
10 PSIG), causing relief
valve to open on jacket. Lower the pressure and
restart sterilization cycle.
GENERAL:
Touchscreen is not responding.
•
Calibrate
touchscreen.
Pressure builds in vessel and
flow goes to Zero.
•
Exhaust filter is blocked. Replace filter and verify that
exhaust heater is on.
Summary of Contents for BioFlo 415
Page 12: ......
Page 129: ...117 New Brunswick Scientific User s Guide Figure 65a Controller Schematics left side ...
Page 130: ...118 BioFlo 415 y M1360 0050 User s Guide Figure 65a Controller Schematics right side ...
Page 145: ...133 New Brunswick Scientific User s Guide ...
Page 146: ...134 BioFlo 415 y M1360 0050 User s Guide ...
Page 147: ...135 New Brunswick Scientific User s Guide ...
Page 148: ...136 BioFlo 415 y M1360 0050 User s Guide ...
Page 149: ...137 New Brunswick Scientific User s Guide ...
Page 150: ...138 BioFlo 415 y M1360 0050 User s Guide ...
Page 151: ...139 New Brunswick Scientific User s Guide ...
Page 152: ...140 BioFlo 415 y M1360 0050 User s Guide ...
Page 153: ...141 New Brunswick Scientific User s Guide ...
Page 154: ...142 BioFlo 415 y M1360 0050 User s Guide ...
Page 155: ...143 New Brunswick Scientific User s Guide ...
Page 156: ...144 BioFlo 415 y M1360 0050 User s Guide ...
Page 157: ...145 New Brunswick Scientific User s Guide ...