4-2
4.1.2 Manually Discharging a Patient
To manually discharge a patient:
1. Select
the
[
Patient Setup
] QuickKey, or [
Main Menu
][
Patient Setup >>
].
2. Select
[
Discharge Patient
]. In the popup menu, you can either:
Directly select [
Ok
] to discharge the current patient, or
Select [
Standby
] then [
Ok
]. The patient monitor enters the standby mode after
discharging the current patient, or
Select [
Cancel
] to exit without discharging the patient.
4.2 Admitting a Patient
4.2.1 Auto Admitting a Patient
The monitor automatically admits a new patient in the following situations:
After a patient is manually discharged, the monitor automatically admits a new patient.
After being switched off for the selected time period, the monitor automatically
discharges the previous patient and admits a new patient at startup.
Always inputs patient information as soon as the patient is admitted. For details, refer to
4.3.1
Editing Patient Information
.
WARNING
z
The settings of patient category and paced status always contain a default value,
regardless of whether the patient is admitted or not. Check if the setting is correct
for your patient.
z
For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the
patient monitor could mistake a pace pulse for a QRS and fail to alarm when the
ECG signal is too weak.
z
For non-paced patients, you must set [Paced] to [No].
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...