A-5
PACE rejection/enhancement
Pace enhancement
Signal amplitude: Voh2.5V
Pulse width: 10ms±5%
Signal rising and falling time: 100μs
Nurse Call Signal
Amplitude
High level: 3.5 to 5 V, providing a maximum of 10 mA output
current;
Low level: < 0.5 V, receiving a maximum of 5 mA input
current.
Rising and falling time
1 ms
Defib Sync Pulse
Output impedance
100
Max time delay
35 ms (R-wave peak to leading edge of pulse)
Amplitude
High level: 3.5 to 5 V, providing a maximum of 10 mA output
current;
Low level: < 0.5 V, receiving a maximum of 5 mA input
current.
Pulse width
100 ms ±10%
Rising and falling time
1 ms
Alarm output (Network connector)
Alarm delay time from the
patient monitor to remote
equipment
The alarm delay time from the patient monitor to remote
equipment is 2 seconds, measured at the monitor’s signal
output connector.
A.5 Data Storage
Trends
Trends: 120 hours, at 1 min resolution
Mid-length trends: 4 hours, at 5 s resolution
Long-trends: 1200 hours, at 10 min resolution
Parameter alarms
1800 physiological alarms and manual events and related
parameter waveforms.
Arrh. events
128 arrhythmia events and relate waveforms and parameters.
NIBP measurements
1600 sets
Full-disclosure waveforms
48 hours at maximum. The specific storage time depends on
the waveforms stored and the number of stored waveforms.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...