WARNING
z
If measuring intracranial pressure (ICP) with a sitting patient, level the transducer
with the top of the patient’s ear. Incorrect leveling may give incorrect values.
14.5 Understanding the IBP Display
The IBP measurement is displayed on the monitor as a waveform and numeric pressures. The
figure below shows the waveform and numerics for the Art pressure. For different pressures,
this display may be slightly different.
(1)
Waveform
(2) Pressure
unit
(3) Systolic
pressure
(4) Diastolic
pressure
(5) Mean
pressure
For some pressures, the parameter window may show the mean pressure only. For different
pressures, their defaults unit may be different. If the Art and ICP pressures are measured
simultaneously, the ICP parameter area will display numeric CPP, which is obtained by
subtracting ICP from the Art mean.
(1)
(3) (5)
(4)
(2)
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...