7-5
7.4 Alarm Tone Configuration
7.4.1 Setting the Minimum Alarm Volume
1. Select
[
Main Menu
][
Maintenance >>
][
User Maintenance >>
]enter the
required password.
2. Select
[
Alarm Setup >>
] to enter the [
Alarm Setup
] menu.
3. Select
[
Minimum Alarm Volume
] and toggle between 0 and 10.
The minimum alarm volume refers to the minimum value you can set for the alarm volume,
which is not affected by user or factory default configurations. The setting of minimum
alarm volume remains unchanged when the patient monitor shuts down and restarts.
7.4.2 Changing the Alarm Volume
1. Select
the
[
Volume Setup
] QuickKey or the [
Alarm Setup
] QuickKey
ė
[
Others
], or
[
Main Menu
][
Alarm Setup >>
]
ė
[
Others
].
2.
Select the appropriate volume from [
Alm Volume
]: X-10, in which X is the minimum
volume, depending on the set minimum alarm volume, and 10 the maximum volume.
3. Select
[
High Alarm Volume
] to set the volume of the high priority alarm as [
Alm
0
], [
Alm 1
] or [
Alm 2
].
4. Select
[
Reminder Vol
] to set the volume of the reminder tone as
[High
], [
Med
] or
[
Low
].
When alarm volume is set to 0, the alarm sound is turned off and a
symbol appears on
the screen.
7.4.3 Setting the Interval between Alarm Sounds
You cannot change the interval between alarm tones if you choose mode 1 or 2 as your
desired alarm tone pattern. For these two patterns, the interval between alarm tones
identifies the alarm levels as follows:
Mode 1:
Interval between high level alarm tones:
continuously.
Interval between medium level alarm tones:
5 s.
Interval between low level alarm tones:
20 s.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...