3-13
3.11.4 Demo Mode
In Demo mode, the monitor can demonstrate its major functions when patient or patient
simulator is not connected. The Demo mode is password protected.
To enter the Demo mode:
1. Select
[
Main Menu
][
Maintenance >>
].
2. Select
[
Demo >>
]. Enter the required password and then select [
Ok
].
To exit the Demo mode, select [
Main Menu
][
Maintenance >>
][
Exit Demo
].
WARNING
z
The Demo mode is for demonstration purpose only. To avoid that the simulated
data are mistaken for the monitored patient’s data, you must not change into
Demo mode during monitoring. Otherwise, improper patient monitoring and
delayed treatment could result.
3.11.5 Standby Mode
In standby mode, you can temperately stops patient monitoring without turning off the
monitor. To enter the standby mode, select the Standby QuickKey
.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...