22-3
22.4 Nurse Call
The patient monitor provides a nurse call connector to output nurse call signal when a
userdefined alarm occurs. To obtain nurse call signal, use the nurse call cable (
PN:
009-003116-00
) we supply to connect the hospital nurse call system to the nurse call
connector of the monitor and then follow this procedure:
1. Select
[
Main Menu
][
Maintenance >>
][
User Maintenance >>
]enter the
required password.
2. Select
[
Others >>
] to access the [
Others
] menu.
3. Select
[
Nurse Call Setup >>
] to change the nurse call settings as follows:
Select [
Signal Type
] and toggle between [
Pulse
] and [
Continuous
].
[
Pulse
]: the nurse call signals are pulse signals and each pulse lasts 1 second. When
multiple alarms occur simultaneously, only one pulse signal is outputted. If an
alarm occurs but the previous one is not cleared yet, a new pulse signal will also be
outputted.
[
Continuous
]: the nurse call signal lasts until the alarm ends, i.e. the duration of a
nurse call signal equals to that of the alarm condition.
Select [
Contact Type
] and toggle between [
Normally Open
] and [
Normally Closed
].
[
Normally Open
]: select if your hospital’s nurse call relay contact is normally
open.
[
Normally Closed
]: select if your hospital’s nurse call relay contact is normally
closed.
Select [
Alm Lev
] and set the alarm level for nurse call-triggering alarms.
Select [
Alarm Cat
.] and then select the category to which the nurse call-triggering
alarms belong.
Alarm conditions are indicated to nurses only when:
The nurse call system is enabled,
An alarm that meets your preset requirements occurs, and
The monitor is not in the alarm paused or reset status.
WARNING
z
To obtain the nurse call signal, use the nurse call cable (
PN: 009-003116-00
) we
supply. Otherwise the nurse call function will not work and the monitor may be
damaged.
z
Do not rely exclusively on the nurse call system for alarm notification. Remember
that the most reliable alarm notification combines audible and visual alarm
indications with the patient’s clinical condition.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...