D-4
Measurement Alarm message
L
I
Cause and solution
ECG Low Freq. Noise
L
A
Low frequency signals are detected on the
ECG analysis lead. Check for any possible
source of interference around the cable and
electrode.
ECG Amplitude Too
Small
L
C
The ECG amplitude didn’t reach the
detected threshold. Check for any possible
source of interference around the cable and
electrode.
ECG Config. Err
L
C
ECG configuration is wrongly downloaded.
Check the downloaded configuration and
re-download the correct configuration.
Resp
Resp Disturbed
L
A
The respiration circuit is disturbed. Restart
the monitor.
Temp
Temp Cal. Err
H
C
A calibration failed. Restart the monitor.
T1 Sensor Off
L
A
The Temp sensor has become detached from
the patient or the module. Check the sensor
connections.
T2 Sensor Off
L
A
SpO
2
SpO
2
Sensor Off
L*
B
The SpO
2
sensor has become detached from
the patient or the module, or there is a fault
with the SpO
2
sensor, or an unspecified
SpO
2
sensor has been used. Check the
sensor application site and the sensor type,
and make sure if the sensor is damaged.
Reconnect the sensor or use a new sensor.
SpO
2
Sensor Fault
L
C
SpO
2
No Sensor
L
B
SpO
2
Unknown
Sensor
L C
SpO
2
Sensor
Incompatible
L C
SpO
2
Too Much Light
L
C
There is too much light on the SpO
2
sensor.
Move the sensor to a place with lower level
of ambient light or cover the sensor to
minimize the ambient light.
SpO
2
Low Signal
L
C
The SpO
2
signal is too low or too weak.
Check the patient’s condition and change
the sensor application site. If the error
persists, replace the sensor.
SpO
2
Weak Pulse
L
C
SpO
2
Interference
L
C
The SpO2 signal has been interfered. Check
for any possible sources of signal noise
around the sensor and check the patient for
great motion.
SpO
2
Board Fault
L
C
There is a problem with the SpO
2
measurement board. Do not use the module
and contact your service personnel.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...