14-1
14
Monitoring IBP
(for uMEC12/uMEC15)
14.1 Introduction
The monitor can monitor up to 2 invasive blood pressures and displays the systolic, diastolic
and mean pressures and a waveform for each pressure.
14.2 Safety
WARNING
z
Use only pressure transducers specified in this manual. Never reuse disposable
pressure transducers.
z
Make sure that the applied parts never contact other conductive parts.
z
To reduce the hazard of burns during high-frequency surgical procedure, ensure
that the monitor’s cables and transducers never come into contact with the
high-frequency surgical units.
z
When using accessories, their operating temperature should be taken into
consideration. For details, refer to instructions for use of accessories.
z
The neutral electrode of the electro-surgery unit (ESU) shall properly contact the
patient. Otherwise, burns may result.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...