8-14
8.6.6 Deleting a Reference Segment
To delete the current ST reference segment, select [
Delete Ref.
] in the [
ST Analysis
] menu
and then select [
Ok
] in the popup.
8.6.7 Recording the ST Segment
To record the current ST segment and reference segment, select [
Record
] in the [
ST Analysis
]
menu.
8.6.8 Changing the ST Alarm Limits
High and low ST alarm limits can be set individually for each ECG lead. Alarm limits can
also be set separately for single-lead and multi-lead ST monitoring. You can select [
ST
Alarm Setup >>
] from [
ST Analysis
] menu and then change ST alarm settings for each lead.
8.6.9 Setting the ST Alarm Delay Time
To set the ST alarm delay time,
1. Select
[
Main Menu
][
Maintenance >>
][
User Maintenance >>
]. Enter the
required password and then select [
OK
].
2. Select
[
Alarm Setup >>
][
ST Alarm Delay
].
8.6.10 Adjusting ST Measurement Points
As shown in the figure below, the ST measured for each beat complex is the vertical
difference between two measurement points with the R-wave peak as the baseline for the
measurement.
R-wave peak
P
T
Isoelectric point
ST measurement point
Difference=ST value
J point
Q S
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...