15-3
NOTE
z
The above picture is connecting illustration when TI sensor PN 6000-10-02079 is
used. The connection may be different if other TI sensors are used.
3.
Select the C.O. parameter window to enter the [
C.O. Setup
] menu. Check if the height
and weight are appropriate for your patient. Change if necessary.
4. In
the
[
C.O. Setup
] menu:
Check that the correct computation constant is entered. Refer to the Instruction for
Use of pulmonary artery catheter to determine the [
Comp. Const
] according to the
entered injectate volume and temperature. To change the computation constant,
select [
Comp. Const
] and then enter the correct value. When a new catheter is used,
the computation constant should be adjusted in accordance with the manufacturer’s
instructions for use.
Set the [
Auto TI
] to [
Manual
] or [
Auto
]. If you select [
Auto
], the system
automatically detects the injectate temperature, and the [
Manual TI
] is disabled. If
you select [
Manual
], you need to enter the injectate temperature at [
Manual TI
]
manually.
Set the [
Measuring Mode
] to [
Manual
] or [
Auto
]. In [
Auto
] mode, the monitor
automatically takes C.O. measurement after establishing a baseline blood
Patient monitor
C.O. cable
Injectate
Syringe
In-line probe
PA catheter
Balloon inflation port
Ice water
Stopcock
Hole
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...