E-4
E.6 Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All
measurements have a true RMS only
The following outlet conditions apply when performing the Patient Leakage Current test.
normal polarity (Normal Condition)
reverse polarity (Normal Condition),
normal polarity with open neutral (Single Fault Condition)
reverse polarity with open neutral (Single Fault Condition)
normal polarity with open earth (Single Fault Condition)
reverse polarity with open earth (Single Fault Condition)
LIMITS
For CF
applied parts
10 A in Normal Condition
50 A in Single Fault Condition
For BF
applied parts
100 A in Normal Condition
500 A in Single Fault Condition
E.7 Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage,
through a limiting resistance, to selected applied part terminals. Current measurements are
then taken between the selected applied part and earth. Measurements are taken with the
test voltage (110% of mains) to applied parts in the normal and reverse polarity conditions
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal Polarity
Reversed Polarity
LIMITS
For CF
applied parts: 50 A
For BF
applied parts: 5000 A
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...