A-7
Defibrillation Protection
Enduring 5000V (360 J) charge without data loss or corruption
Baseline recovery time: <5 s (after defibrillation)
Polarization recovery time: <10 s
Defibrillation energy absorption: <10% (100 load)
Calibration signal
1mV (peak-to-peak value)
Accuracy:
±
5
%
ESU protection
Cut mode: 300 W
Coagulate mode: 100 W
Recovery time: 10 s
In compliance with the requirements in clause 202.6.2.101 of
IEC 60601-2-27
Pace Pulse
Pace pulse markers
Pace pulses meeting the following conditions are labelled with a
PACE marker:
Amplitude:
Width:
Rise time:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 μs
Pace pulse rejection
When tested in accordance with the IEC60601-2-27:
201.12.1.101.13, the heart rate meter rejects all pulses meeting
the following conditions.
Amplitude:
Width:
Rise time:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 μs
HR
Measurement range
Neonate:
Pediatric:
Adult:
15 to 350 bpm
15 to 350 bpm
15 to 300 bpm
Resolution 1
bpm
Accuracy
±1 bpm or ±1%, whichever is greater.
Sensitivity
200 V (lead II)
HR averaging method
In compliance with the requirements in Clause 201.7.9.2.9.101
b) 3) of IEC60601-2-27, the following method is used:
If the last 3 consecutive RR intervals are greater than 1200 ms,
the 4 most recent RR intervals are averaged to compute the HR.
Otherwise, heart rate is computed by subtracting the maximum
and minimum ones from the most recent 12 RR intervals and
then averaging them.
The HR value displayed on the monitor screen is updated every
second.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...