B-1
B
EMC and Radio Regulatory Compliance
B.1 EMC
The device meets the requirements of IEC 60601-1-2.
NOTE
z
Using accessories, transducers and cables other than those specified may result in
increased electromagnetic emission or decreased electromagnetic immunity of the
patient monitoring equipment.
z
The device or its components should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the device or its components
should be observed to verify normal operation in the configuration in which it will
be used.
z
The device needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided below.
z
Other devices may interfere with this device even though they meet the
requirements of CISPR.
z
When the inputted signal is below the minimum amplitude provided in technical
specifications, erroneous measurements could result.
z
Portable and mobile communication equipment may affect the performance of this
monitor.
z
Other devices that have RF transmitter or source may affect this device (e.g. cell
phones, PADs, PCs with wireless function).
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...