A-6
A.6 Wireless Network
Standards
MSD45N Wireless Module: IEEE 802.11a/b/g/n, support Wi-Fi
A.7 Measurement Specifications
The adjustable range of alarm limits is the same with the measurement range of signals
unless otherwise specified.
A.7.1 ECG
ECG
Standards IEC60601-2-27
Lead set
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
12-lead: I, II, III, aVR, aVL, aVF, V1 to V6 (for
uMEC12/uMEC15)
ECG standard
AHA, IEC
Display sensitivity
1.25 mm/mV (X0.125), 2.5 mm/mV (X0.25), 5 mm/mV (X0.5),
10 mm/mV (X1), 20 mm/mV (X2), 40 mm/mV (X4)
Accuracy: ±5%
Sweep speed
6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Accuracy:
±
5
%
Bandwidth (-3dB)
Diagnostic mode:
Monitor mode:
Surgical mode:
ST mode:
0.05 to 150 Hz
0.5 to 40 Hz
1 to 20 Hz
0.05 to 40 Hz
Common mode rejection ratio
(with Notch off)
Diagnostic mode:
Monitor mode:
Surgical mode:
ST mode:
90 dB
105 dB
105 dB
>105 dB(with Notch on)
Notch on/off
Differential input impedance
5 M
Input signal range
±8 mV (peak-to-peak value)
Electrode offset potential
tolerance
±500 mV
Lead-off detection current
Measuring electrode: <0.1 A
Drive electrode: <1 A
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...