D-2
Measurement Alarm messages
L
Cause and solution
Extreme Brady
H
R on T
M*
Run PVCs
L*
PVCs/min
M*
Multif.PVC
M*
Bigeminy
M*
Trigeminy
M*
Tachy
M*
Brady
M*
Vent. Rhythm
M*
Multif. PVC
M*
Nonsus. Vtac
M*
Pause
L*
Resp
Resp Apnea
H
The respiration signal was so weak that the
monitor cannot perform respiration analysis.
Check the patient’s condition and the Resp
connections.
Resp Artifact
H
The patient’s heartbeat has interfered with
his respiration. Check the patient’s condition
and the Resp connections.
SpO
2
SpO
2
Desat
H
The SpO2 value has fallen below the
desaturation alarm limit. Check the patient’s
condition and check if the alarm limit
settings are correct.
No Pulse
H
The pulse signal was so weak that the
monitor cannot perform pulse analysis.
Check the patient’s condition, SpO
2
sensor
and measurement site.
CO
2
CO
2
Apnea
H
The patient stops breathing, or the
respiration signal was so weak that the
monitor cannot perform respiration analysis.
Check the patient’s condition and the RM
connections.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...