8-19
Arrh. event
Range
Default
Step
Unit
Vtac Rate
100 to 200
Adult, pediatric: 130
Neonate: 160
5 bpm
Vtac PVCs
3 to 99
6
1
/min
Pause Time
1.5, 2.0,2.5
2
/
s
Vbrd PVCs
3 to 99
5
1
/min
Vbrd Rate
15 to 60
40
5
bpm
8.7.4 Setting the Extended Arrhythmia
The following arrhythmia events are defined as extended arrhythmia:
Extreme Tachy
Extreme Brady
Vent. Brady
Nonsus. Vtac
Multif. PVC
Irr. Rhythm
Pause
Afib
You can select [
Main Menu
][
Maintenance >>
][
User Maintenance >>
]enter the
required passwordselect [
Alarm Setup >>
], and set [
Extended Arrh.
] to [
Enable
] or
[
Disable
]. When [
Extended Arrh.
] is set to [
Disable
], the patient monitor does not analysis
the extended arrhythmia events and corresponding alarms are not given.
CAUTION
z
Set [
Extended Arrh.
] to [
Disable
] when the patient monitor is connected to the
Central Monitoring System of version prior to 06.01.00. Failure to do so may cause
the Central Monitoring System unable to display extended arrhythmia related
alarms normally when extended arrhythmia occurs.
8.7.5 Reviewing Arrhythmia Events
Please refer to
18
Review
.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...