11-6
Symptoms
Possible Cause
Correction Action
Dashes “- -” display in
place of numerics.
Measurement is invalid.
Check that the sensor is properly
applied. Change the application site if
necessary.
Do not see SpO
2
parameter tiles in
display.
Parameter not configured to
display.
Switch the SpO
2
monitoring function on
as described in
3.10.1 Switching the
Parameters On/Off
.
Unable to obtain SpO
2
reading
Patient has poor perfusion
Switch limbs/notify physician
Sensor not on patient
Check if the “SPO2 Sensor Off” alarm
is reported.
If so, reapply the sensor.
If not, contact the service personnel.
Cables loose/not connected
Check connections, switch cable
Ambient light
Check if the “SpO2 Too Much Light”
alarm is reported. If so, move the sensor
to a place with lower level of ambient
light or cover the sensor to minimize the
ambient light.
No SpO2 waveform
Waveform not selected to
display
Switch the SpO
2
monitoring function on
as described in
3.10.1 Switching the
Parameters On/Off
.
Cable or sensor not plugged
in
Check that the cable is properly
connected and sensor securely applied.
Low amplitude SpO
2
signal
SpO
2
sensor on same limb
as cuff
Check that the sensor is properly
applied. Change the application site if
necessary.
Patient has poor perfusion
Change the application site.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...