15-1
15
Monitoring Cardiac Output
(for uMEC12/uMEC15)
15.1 Introduction
The cardiac output (C.O.) measurement invasively measures cardiac output and other
hemodynamic parameters using the right heart (atria) thermodilution method. A cold solution
of known volume and temperature is injected into the right atrium through the proximal port
of a pulmonary artery (PA) catheter. The cold solution mixes with the blood in the right
ventricle and the change in blood temperature is measured with a thermistor at the distal end
of the catheter in the pulmonary artery. The temperature change is displayed as a curve in the
C.O. split screen, and the monitor calculates the C.O. value from this curve. The C.O. value
is inversely proportional to the area under the curve. As cardiac output varies continuously, a
series of measurements must be carried out to achieve a reliable C.O. average value. Always
use the average of multiple thermodilution measurements for therapy decisions. The monitor
is capable of storing 6 measurements.
WARNING
z
C.O. monitoring is restricted to adult patients only.
15.2 Understanding the C.O. Display
The C.O. measurement is displayed on the monitor as numeric C.O., C.I. and TB in the C.O.
parameter window as shown below. To enter the [
C.O. Setup
] menu, select the C.O.
parameter window.
(1) Cardiac
output
(2) Time at which the C.O. average is calculated
(3) Cardiac
index
(4) Blood
temperature
(1)
(3)
(4)
(2)
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...