A-14
A.7.7 IBP (for uMEC12/uMEC15)
Standards
Meet standard of IEC60601-2-34.
Technique Direct
invasive
measurement
IBP
Measurement range
-50 to 300 mmHg
Resolution 1
mmHg
Accuracy
±2% or ±1 mmHg, whichever is greater (without sensor)
Refreshing rate
1 s
Pressure transducer
Excitement voltage
5 VDC, ±2%
Sensitivity 5
V/V/mmHg
Zero adjustment range
f
200 mmHg
Impedance range
300 to 3000
Alarm limit
Range (mmHg)
Step (mmHg)
Sys High
(low limit + 2) to 300
1
Mean High
Dia High
Sys Low
-50 to (high limit – 2)
Mean Low
Dia Low
A.7.8 C.O. (for uMEC12/uMEC15)
Measurement method
Thermodilution method
Measurement range
C.O.:
0.1 to 20 L/min
TB:
23 to 43
ć
TI:
0 to 27
ć
Resolution
C.O.:
TB, TI:
0.1 L/min
0.1
ć
Accuracy
C.O.:
TB, TI:
±5% or ±0.1 L /min, whichever is greater
±0.1
ć
(without sensor)
Repeatability
C.O.:
±2% or ±0.1 L/min, whichever is greater
Alarm range
TB:
23 to 43
ć
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...