D-8
Measurement Alarm message
L
I
Cause and solution
Recorder Vlt Low
L
C
supply. Restart the monitor.
Recorder Head Hot
L
C
The recorder has been working for too long
time. Stop the recording and resume the
recording till the recorder’s printhead cools
down.
Rec Paper Wrong Pos.
L
A
Re-load the recorder paper.
System
System Watchdog Err
H
C
An error occurred to the system. Restart the
monitor.
System Software Err
H
C
System CMOS Full
H
C
System CMOS Err
H
C
System FPGA Err
H
C
System Err N
H
C
N is within 2 to 12.
Other Bed
Disconnected
L A
Check
network
connection.
PWR interrupted.
Check meas. state
L
A
Power supply failed accidently. Check the
measurements when the monitor restarts.
No CMS
L
A
The monitor is disconnected from the CMS.
Check network connection.
Summary of Contents for uMEC10
Page 1: ...uMEC10 uMEC12 uMEC15 Patient Monitor Operator s Manual ...
Page 2: ......
Page 24: ...1 6 FOR YOUR NOTES ...
Page 29: ...2 5 2 3 2 uMEC12 uMEC15 1 6 7 2 8 9 10 3 4 5 ...
Page 36: ...2 12 FOR YOUR NOTES ...
Page 50: ...3 14 FOR YOUR NOTES ...
Page 58: ...4 8 FOR YOUR NOTES ...
Page 72: ...6 8 FOR YOUR NOTES ...
Page 118: ...8 28 FOR YOUR NOTES ...
Page 174: ...18 8 FOR YOUR NOTES ...
Page 186: ...20 6 FOR YOUR NOTES ...
Page 192: ...21 6 FOR YOUR NOTES ...
Page 202: ...23 4 FOR YOUR NOTES ...
Page 206: ...24 4 FOR YOUR NOTES ...
Page 224: ...26 10 FOR YOUR NOTES ...
Page 248: ...B 8 FOR YOUR NOTES ...
Page 266: ...C 18 FOR YOUR NOTES ...
Page 280: ...E 6 FOR YOUR NOTES ...
Page 286: ...F 6 FOR YOUR NOTES ...
Page 287: ......
Page 288: ...PN 046 008784 00 1 0 ...